1. Measuring range: residual chlorine: 0-20.00mg/L, resolution: 0.01mg/L;
|
HOCL: 0-10.00mg/L resolution: 0.01mg/L;
|
2. Temperature: 0- 99.9 ℃, resolution: 0.1 ℃
|
3. Accuracy: Residual chlorine: ± 2% or ± 0.035mg / L, take the bigger;
|
4. HOCL: ± 2% or ± 0.035mg / L, take the bigger;
|
5. Temperature: ± 0.5 ℃ (0 ~ 60.0 ℃);
|
6. Sample temperature: 0 ~ 60.0 ℃, 0.6MPa;
|
7. Sample flow rate: 200 ~ 250 mL/1min Automatic and Adjustable
|
8. Minimum detection limit: 0.01mg / L; Stability: ± 0.02 (mg / L)/24h
|
9. Isolated current output: 4~20 mA(load <750Ω)
|
10. High and low alarm relays: AC220V, 7A;
|
Hysteresis Amount 0- 5.00mg / L, Arbitrary Regulation
|
11.Communication interface:RS485(optional)
|
12. Power Supply: AC220V ± 22V, 50Hz ± 1Hz; DC24V (optional).
|
13. Overall dimension: 96 (length) x 96 (width) x 110 (depth) mm
|
dimension of the hole: 92 x 92mm
|
14. Power: ≤ 10W
|
15. Working Conditions: ambient temperature: 0 ~ 60 ℃; relative humidity <85%;
|
16. Weight: 0.8kg
|
17. Adopt the flow-through installation, and incoming and outgoing water mouth diameter at Φ10.
|
Note: The user can select two-way current output; one way of residual chlorine or HOCL corresponds to current output, and the other way of optional temperature corresponds to the current output. The ex-factory standard configuration is that one way of residual chlorine corresponds to the current output. Please specify when ordering.
|