1. Measuring range
|
Residual chlorine: 0-20.00mg/L, Resolution: 0.01mg/L;
|
HOCL: 0-10.00mg/L Resolution: 0.01mg/L;
|
pH value: 0 - 14.00pH Resolution: 0.01pH;
|
Temperature: 0- 99.9 ℃ Resolution: 0.1 ℃
|
2. Accuracy: Residual chlorine: ± 2% or ± 0.035mg / L, take the bigger;
|
HOCL: ± 2% or ± 0.035mg / L, take the bigger;
|
pH value: ± 0.05Ph
|
Temperature: ± 0.5 ℃ (0 ~ 60.0 ℃);
|
3. Sample temperature: 0 ~ 60.0 ℃, 0.6MPa;
|
4. Sample flow rate: 200 ~250 mL/1min Automatic and Adjustable
|
5. Minimum detection limit: 0.01mg / L
|
6. Isolated current output: 4~20 mA(load <750Ω)
|
7. High and low alarm relays: AC220V, 7A;
|
hysteresis 0- 5.00mg / L, arbitrary regulation
|
8. RS485 communication interface (optional)
|
It can be convenient to computer monitoring and communication
|
9. The data storage capacity: 1 month (1 point/5 minutes)
|
10. Power Supply: AC220V ± 22V, 50Hz ± 1Hz; DC24V (optional).
|
11. Protection grade: IP65
|
12. Overall dimension: 146 (length) x 146 (width) x 108 (depth) mm;
|
dimension of the hole: 138 x 138mm
|
Note: The wall installation may be ok, please specify when ordering.
|
13. Weight: Secondary Instrument: 0.8kg, flow cell with residual chlorine,
|
pH electrode weight: 2.5kg;
|
14. Working Conditions: ambient temperature: 0 ~ 60 ℃; relative humidity <85%;
|
15.Adopt the flow-through installation, inlet and outlet diameter at Φ10.
|
Note: The user can select two-way current output; one way of residual chlorine corresponds to current output, and the other way of optional pH or temperature corresponds to the current output. The ex-factory standard configuration is that one way of residual chlorine corresponds to the current output. Please specify when ordering
|